Others

<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
<?=$ecms_gr[title]?>
WM-IP02 Infusion Pump Technical Specification
 
Infusion Mode Volume Mode, Time Mode, Drug Mode
Rate Unit ml/h, volume/time, d/min
IV Sets Standard single use IV sets ( 10,15,20,60d/mL, with diameter 3.4 ~ 4.5mm)
Flow Rate 1.0ml/h~1200ml/h
1.0d/min~400d/min
Increment 1.0-99.9ml/h in 0.1ml/h step;100-1200ml/ in 1ml/h step
1-400d/min in 1d/min step;
Deviation Flow Rate:≤5%(after calibration)
VTBI 0.1~9999.9 ml
Total Volume 0.0~9999.9 ml
Time 1min~99h00min
Purge Rate 600ml/h,display purge volume
KVO 1.0ml/h~5.0ml/h,adjustable
Keypad Numeric keypad,
BOLUS Bolus function available, flow rate adjustable
Night Mode Screen to be dim automatically in 5mins after standby time expired.
Occlusion pressure detector downstream pressure detection, 40KPa~130KPa, 10 levels adjustable with dynamic
pressure value display
Air Bubble Detector Ultrasonic detection, ≥50ul,Single bubble volume: 50~300 μl, 10 levels adjustable.
Volume 10 levels adjustable
Motor Anti-reverse Function Anti-reverse function to prevent upstream
IV Set Management Open system, accept different IV brands after calibration
Self-Testing Self-Testing before startup and during infusion
Audio Visual Alarm Message Air bubble, downstream occlusion, door open, infusion finish, standby time expired, battery low, battery empty, battery disengaged, AC Power disengaged, motor error, pump error, pressure sensor error, parameter error, system error
 
Screen 3.2inch brightness LCD
Power Supply AC Power:100V~240V,50Hz/60Hz,≤25VA
Battery Rechargeable Lithium battery, DC11.1V,3000mAh;operating time: ≥10h @25ml/h, displaying remaining battery
Operation Condition Temperature:+5℃~+40℃, Humidity:20%~90%,Air pressure:70~106KPa
Transport & Storage Temperature :-20℃~+55℃,Humidity:10%~95%,Air pressure:50~106KPa
Classification Type CF, Class I, IPX3
Dimension 145mm×150mm×200mm;Weight:1.7kg